3 - 31
3.5 Operation for Image Observation
Follow the operation below for observation of a scanning image.
3.5.1 Selecting a Detector
A detector can be selected by using the DETECTOR block located in the Image tab on the
Operation Panel. One of the following detectors can be selected: a secondary electron detector
(SE), a backscattered electron detector (BSE), or (optional) other detectors. If the ABCC Link
is checked, changing detector settings automatically causes contrast and brightness
adjustments.
Fig. 3.5-1 Selecting a Detector
HINT: Changing detectors causes a substantial change in the brightness of the image.
If the ABCC Link check box in the Image tab of the Operational panel has been checked,
ABCC will start automatically when detector selection is changed.
Image brightness will be adjusted to an adequate value.