For Safe Handling
SAFETY - 17
Precautions on Proper Operation (cont’d)
OS Operation during S-3400N SEM Operation
Although the Windows Task Manager can be launched from the
Taskbar or by Ctrl+Alt+Del key operation, if Standby or Halt mode is
set or logged-off while the S-3400N control program is running, it will
not work normally when recovered from such a status. Close
S-3400N control program if setting of such mode is necessary.
USB Devices
When connecting USB devices such as external storage unit, printer
etc., take notice of the following.
(1) USB Port 1 is used by the display control unit, ports available to
users are subject to restrictions.
(2) Terminate S-3400N control program when carrying out disk
formatting or copying a large volume of image files to MO disk,
hard disk or other storage devices using USB port.
(3) By the above operation, sometimes the message “USB not
connected” will be shown. In such case, S-3400N control program
will not continue normal operation. If this message appears,
close Windows, shut down the PC, turn off the key switch for the
system unit, wait 30 seconds, and then restart the system by
turning on the power again.