Hitachi S-3400N Microscope & Magnifier User Manual


 
3.6.4
3 - 103
Fig. 3.6-6 Settings Window Record Tab
In the above case, resulting file names are [S3400_i001.TIFF], [S3400_i002.TIFF] and so on.
If the Auto Increment is not set when the Data Number Save is selected, Following
message will be shown.
Fig. 3.6-7 Data No. Save Window
If the function is reset so that the above auto incrementing is not performed after the "Save
Data No." option was enabled, the image will be saved with no Data No. (using a specified
file name only). If an attempt is made to save multiple images in a single operation, the
second and subsequent images will be saved under the same name, in which case a
confirmation message will appear, asking the user whether files can be overwritten. In this
case, temporarily cancel the image save, cancel the "Save Data No." option, and save the
images one by one. The captured image is assigned a data number that was in effect
when the capturing process was executed. Therefore, before performing a capturing
process, the auto incrementing option should be enabled.