3.8 Shutting Down
3 - 112
3.8 Shutting Down
3.8.1 Turning High Voltage Off
(1) Close opening dialog windows. Save data if necessary.
(2) Click the Acceleration Voltage OFF button
on the Control Panel to turn off the
acceleration voltage.
3.8.2 Setting the Stage at the Specimen Exchange Position
(1) Type I (manual stage)
Align the axial scale with the specimen change position.
Specimen exchange position X: 30mm Y: 20mm Z: 35mm
R: arbitrary T: 0°
(2) Type II (motor drive stage)
On the right edge of the Control Panel, click either the Change Z65 [HOME Z65] button or
the Change Z10 [HOME Z10] button. Wait until the display window to the left of the button
lights up in blue.
Fig. 3.8-1 Moving to the Specimen Exchange Position
Specimen exchange position X: 60 mm Y: 25 mm R: 0° T: 0°
Z: 10 mm (Change Z10, HOME Z10)
Z: 65 mm (Change Z65, HOME Z65)