(2) Loading a Set of Operating Conditions
Use the following steps to load a set of operating conditions.
(a) Open the Setup dialog window
Select Condition Load from the Operate menu, or
click the Setup button in the tool button section of the
Operation panel and select Op.Cond tab.
Select Load Memorized Condition on it.
(b) Select a condition name
Memorized condition names are listed. Select a name
by clicking it.
Use sample picture and comment for reference of
selection.
(c) Select items to be loaded
Check items ( Vacc to Contrast) you wish to load and
reproduce. Stage position is effective only for Type 2
instrument.
(d) Reproduce column condition
Click the Load button. Memorized data are retrieved
and the electron optical column is set.
If the Stigma/Alignment is included, a message will
appear for confirmation. These are possibly changed if
the loaded condition was saved long before.
In such cases, it may be better not to reproduce Stigma and Alignment data. The
recent data is memorized automatically y abnother condition memory function
(the most recent condition for each of Vacc - Condenser lens setting is memorized and
reproduced automatically).
(3) Deleting an operating condition
Select a condition name in the same manner as above, and click the Delete button.
3.9.7 Pseudo Color Display
Use function of SEM Data Manager for pseudo color imaging.
See 3.11.3.7 Contrast Conversion.