Hitachi S-3400N Microscope & Magnifier User Manual


 
2.3.9
2 - 57
(d) Back Ground Image checkbox
When this box is checked, the data display is overlaid on the scanning image. If it is
not checked, the background of the data display area is black.
(e) Embed into Image checkbox
When this box is checked, the auto data display is embedded into the image data when
the image is saved to disk. If it is not checked, only the image data is saved. This
setting is also applied for Print, Copy and PCI Transfer commands.
If the For Photograph is checked, the position of the auto data display in the saved
image is shifted slightly upward to ensure that it is properly framed in photographs.
It is recommended to not check the box if the saved image data will be used on the
computer only.
NOTICE: If all of the above display items (Magnification to Time and Signal Name) are
selected, data display on the photograph may be overlapped because of limited
display space.
In such a case, remove insignificant items.
(2) PHOTO SPEED Window
(a) Direct Photo
Selects a scanning speed for direct photo recording. The number of lines for the
selected scanning speed is shown under the box.
(b) Memory Photo
Selects number of scanning lines for memory photo recording.
Normal: For 640 × 480 pixel image data
960 lines (16/19 s photographing time)
For 1280 × 960 pixel image data
960 lines (16/19 s photographing time)
For 2560 × 1920 pixel image data
1920 lines (32/38 s photographing time)
Enhance: For 640 × 480 pixel image data
960 lines (16/19 s photographing time)
For 1280 × 960 pixel image data
1920 lines (32/38 s photographing time)
For 2560 × 1920 pixel image data
1920 lines (32/38 s photographing time)
(Scanning time is shown as the value at 60 Hz/50 Hz power line frequency.)
(3) BRIGHTNESS/CONTRAST Window
Sets brightness and contrast for photo recording. These are compensation factors for
brightness and contrast adjusted in the scanning image. Adjust these values if a
photograph does not have adequate brightness or contrast.