1Ć71
Learning About the System
Cal
Slope
Ambient Air
11a 2
43
R
S
W
7.3 6.8 Wash Clean
5
CO
HSC
9. A small amount of foam wash is moved to the
measurement module to remove the 7.3 reagent. The
reagent pump then moves three 6.8 reagent pre-segments
and a main segment to the inner sample port. The
pre-segments are used to remove any residual 7.3 reagent
from the pathways.
Cal
Slope
Ambient Air
11a 2
43
R
S
W
••••••••••••••••••••••••
•••••••••••••••••
••
•••••
7.3 6.8 Wash Clean
5
CO
HSC
10. The pre-segments move through the sample port into the
waste.
Cal
Slope
Ambient Air
11a 2
43
R
S
W
••••••••••••••••••••••••
•••••••••••
••••••••••••••••••••
•••
••••••••••••••••••
•
•
•
•••••
7.3 6.8 Wash Clean
5
CO
HSC
11. When the main segment reaches the sample port, the
sample pump starts and moves the segment to the
measurement module until it is detected by fluid detector 1
and 1A. The main segment moves toward the sample port.
The waste pump starts.
(Continued)