Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć46 Rapidlab 800 Operator's Manual
NOTE: The waste bottle is disposable and can be autoclaved before you discard it.
The waste bottle is not reusable after autoclaving. Cover the waste bottle with the
cap provided and discard the bottle according to your laboratory protocol or
infection control policy.
3. Discard the waste bottle contents according to your laboratory biohazard
protocol.
4. Clean the waste outlets:
a. Wipe the waste outlet cover with a lint-free tissue dampened with a 10%
solution of household bleach.
b. Rinse with reagent water.
5. If required, remove the waste outlet cover and clean the waste outlets as shown
in Figure 3-28:
a. Grasp the waste outlet cover and pull it off.
b. Wipe the outlets with a lint-free tissue dampened with a 10% solution of
household bleach.
c. Clean the waste outlet cover with the bleach solution.
d. Rinse the waste outlets and the waste outlet cover with reagent water.
    
 
  
 
6. Reinstall the waste outlet cover:
a. Press the alignment pins on the bottom of the waste outlet cover into the
system holes.
b. Align the alignment pins on the top of the waste outlet cover with the
system holes, and then align the waste outlets with the holes in the cover.
c. Press the cover firmly into place, leaving no gaps
7. Reinstall the waste bottle, ensuring that the latch slides back in place.
8. Ensure that the waste outlet cover fits tightly into the waste bottle opening.