CAUTION:ĄĂDo not overtighten the reservoir cap. Overtightening can deform
the gasket and cause leaks.
e. Reinstall the reservoir cap and hand-tighten.
6. Continue with Reinstalling a Sensor, page 3-89.
    
Materials required:
S reference electrode refill kit
S hex tool
S lint-free tissue
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 7 Stop System and press Enter.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
2. Remove the reference sensor as shown in Figure 3-47:
a. Push up the latches on the measurement module door and lift the door.
b. Push the spring-loaded latch to the right.
c. Grasp the tab on the reference sensor and pull the sensor up and out of the
measurement module.
      
     
        
 
   
  
!