Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
1Ć66
Rapidlab 800 Operator's Manual
Cal
Slope
Ambient Air
11a 2
43
R
S
W
••••••••••••••••••••
••
•••••••••••••••••
••
•••••••••••••••
•••••
7.3 6.8 Wash Clean
5
CO
HSC
6. The main segment moves toward the sample port. When
the main segment reaches the sample port, the sample
pump starts and moves the segment to the measurement
module until it is detected by fluid detector 1 and 1A. The
CO-ox module pump moves the 7.3 reagent to FD5. The
waste pump starts.
Cal
Slope
Ambient Air
11a 2
43
R
S
W
7.3 6.8 Wash Clean
5
CO
HSC
7. The probe extends and the main segment moves until the
trailing edge is at fluid detector 1. The pumps stop and the
analytes are measured. The CO-ox zero calibration is
performed.
Cal
Slope
Ambient Air
11a 2
43
R
S
W
••••••••••
••
••••
7.3 6.8 Wash Clean
5
CO
HSC
8. The sample port and reagent manifold are washed while
the measurement is being made. The probe fully retracts.
The foam wash and vent valves open, allowing the wash
solution to become segmented with air, creating foam
wash. The reagent pump starts and moves foam wash to
the inner sample port. The reagent pump stops, the waste
pump starts, and the foam wash is emptied from the
sample port. This sequence repeats four times and is
followed by a solid wash segment.
(Continued)