Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

   
   
The following table lists the minimum volumes required by each system to analyze
capillary tube samples:
  
   
840
844, 845
090 µL
140 µL
850
854, 855
110 µL
160 µL
860
864, 865
125 µL
175 µL
NOTE: If an 844, 845, 854, 855, 864, or 865 detects insufficient sample in the
measurement module, the system analyzes the sample as a microsample. If you
want the system to analyze a different set of parameters for the sample, press
Change Parameters to select another panel. Refer to Selecting Parameter Panels,
page 2-29, for more information.
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
CAUTION: Insert the fire-polished end of the capillary tube into the sample port to
prevent damage to the capillary seal.
1. Prepare the sample and insert the fire-polished end of the capillary tube into
the sample port, as shown in Figure 2-3.
    
Sample Port
Capillary Sample
!