System Administration 5Ć85
     
Use this procedure to shut down the system before you perform service and to
restart the system when finished.
1. Access the Shutdown screen from the Menu screen:
a. Select
7 System Utilities and press Enter.
b. Select 3 Shutdown and press Enter.
2. Press
Yes.
A message appears on the screen and on the roll printer, prompting you to wait
before you disconnect the power.
CAUTION:ĄĂDo not unplug the system until the following message appears on
the screen:
...synching disks... done
This operating system message indicates that the system has successfully
completed the shutdown procedures. Unplugging the system before this
message appears can damage the system.
3. When you see this message, disconnect the power cord from the power supply.
CAUTION:ĄĂAdhere to the time interval or damage can occur to the system.
4. Wait at least 10 seconds after disconnecting the power cord, then reconnect the
power cord into the power source.
After a few moments, the system starts initializing. When initializing is
complete, a screen for Analyze Mode appears.
In the event of a power failure or a power surge and the system shut down,
perform a two-point calibration when the system completes initializing.

!
!

