Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć102
Rapidlab 800 Operator's Manual
    
Materials required:
S measurement module lamp
1. Shut down the system from the Menu screen:
a. Select
7 System Utilities and press Enter.
b. Select
3 Shutdown and press Enter.
c. Press Yes.
CAUTION: You must wait at least 1 minute before you disconnect the power cord,
and then wait at least 10 seconds before you reconnect the power cord. If you do
not adhere to the time intervals, you can damage the system.
2. Wait until prompted, and then disconnect the power cord from the power
source.
3. Disconnect the power cord from the side panel of the system.
4. Turn the system around so that the rear panel faces you.
5. Push the lamp cover up and remove the cover.
WARNING
Ensure that the lamp has been off for at least 5 minutes to allow
sufficient time for it to cool.
6. Allow the lamp to cool for at least 5 minutes.
7. Remove the connector from the system as shown in Figure 3-70.
    
Bracket
Lamp Cover
Connector
Clamp

 
7
Menu Code
3
!