Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć58 Rapidlab 800 Operator's Manual
3. Deproteinize the sample path:
a. Prepare the deproteinizer as directed on the package.
b. Select 2 Maintenance and press Enter.
c. Select
1 Deproteinize and press Enter.
d. Invert the deproteinizer vial several times to mix.
e. Insert an aspiration adapter into the sample port and insert the other end
into the deproteinizer.
f. Press Analyze.
g. When prompted, remove the adapter.
h. Wait for the deproteinizing cycle to finish.
A message box appears prompting you to condition the sensors.
4. Perform the cleaning cycle:
a. Press
Yes.
b. Insert an aspiration adapter into the sample port and immerse the other end
in the glutaraldehyde solution.
c. Press
Analyze.
d. When prompted, remove the adapter.
e. Wait 5 minutes for the cleaning cycle to finish.
NOTE:ĄĂPress Cancel if you want to stop cleaning.
The cleaning cycle finishes. The system performs an extended wash sequence.
The Calibrate System message box appears at the end of the wash sequence.
5. Press
No.
6. Initiate a wash sequence from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 8 Wash and press Enter.
NOTE:ĄĂAs you install the biosensors, ensure that the biosensors are in the
correct location. Visually verify that you align the contacts on the biosensors
with the contacts in the measurement module.
CAUTION:ĄĂDo not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
7. When the wash sequence finishes, remove the test/blank glucose and lactate
sensors and reinstall the biosensors as described in Replacing the Glucose and
Lactate Biosensors page 3-86.
2
Menu Code
1
2
Menu Code
8
(from the Main Menu)
!