Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
DĆ4 800 Series Operator's Manual
NOTE: Always use serial port 1 for the ticket printer.
3. Connect the 9-pin connector of the interface cable to serial port 1 on the 800
system. Refer to Figure D-1.
4. Connect the 25-pin connector to the printer.
5. Tighten the hold down screws on the connectors.
6. Restore power to the 800 system and to the printer.
7. Refer to Configuring for External Devices in Section 5 to configure the 800
system for the printer and to Selecting Printing Options in Section 5 to select
the ticket printer as the printer used to print reports.
Connecting the Bar Code Scanner
Use this procedure to connect an 800 system to the 800 bar code scanner.
Materials Required:
S bar code scanner
S 800 system to bar code scanner cable
To prevent electrical shock and damage to either system, disconnect
the 800 system from the AC power source before installing the bar code scanner.
Refer to Shutting Down and Restarting the System in Section 5 for the procedure to
disconnect the 800 system.
1. Disconnect the 800 system from the AC power source.
2. Connect the 9-pin connector of the bar code scanner to the bar code scanner
port on the 800 system. Refer to Figure D-1.
3. Tighten the hold down screws on the connectors.
4. Restore power to the 800 system.
5. Refer to Configuring for External Devices in Section 5 to configure the 800
system for the bar code scanner.
Connecting to a Bayer Diagnostics Data Management
System
Use this procedure to connect the 800 system to a Bayer Diagnostics data
management system.
Materials Required:
S data management to 800 system to interface cable (supplied with the data
management system)
WARNING