Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
    
Performing the Automatic Clean Sequence
Use this procedure to manually initiate the cleaning sequence. The system
automatically initiates the cleaning sequence every 24 hours at 02:00.
1. Access Auto Clean from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 6 Auto Clean and press Enter.
NOTE:ĄĂTo interrupt the Auto Clean sequence at any time, press Cancel.
The Auto Clean sequence takes 10 minutes. A message box appears prompting
you to perform a two-point calibration.
2. Press
Yes to perform a two-point calibration.
Performing a Wash
1. Access wash from the Menu screen:
a. Select 2 Maintenance and press Enter.
b. Select
8 Wash and press Enter.
2. Press
Home to return to the Ready screen.
Priming the System
1. Access the Prime screen from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 3 Prime and press Enter.
2. Select the appropriate reagent to prime and press
Enter.
3. Press
Done.
4. Press
Home or Menu when the Prime menu appears.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
5. Take the appropriate action.
If . . . Then . . .
you want to perform a
two-point calibration
press Yes.
you do not want to initiate a
two-point calibration
press No.
press
Home to return to the Ready screen.
2
Menu Code
6
2
Menu Code
8
2
Menu Code
3