If the patient ID entered at the 800 system and the 270 CO-oximeter do not match,
the Patient ID Mismatch message box appears.
If . . . Then . . .
you want to combine
the 270 results with
the 800 system results
a. Press Combine Results. The Patient Information screen
appears.
b. Type the patient ID and press
Done.
you do not want to
combine the 270
results with the 800
system results
press Cancel.
Separate reports are generated at the 270 COĆoximeter
and at the 800 system.
If the 800 system completes analysis before the 270 CO-oximeter completes
analysis, the Waiting for CO-oximetry Results message box appears.
If . . . Then . . .
you want to combine
results
press Combine Results. The 800 system creates a combined
report when the results are ready.
you do not want to
combine results
press Cancel. Separate reports will be generated at the 270
CO-oximeter and at the 800 system.