5. Perform the appropriate action.
test another reagent repeat steps 2 through 4.
return to the Menu
screen
press Exit Test.
6. Press Exit Menu.
Use this procedure to test the flow rate of the reagent pump and to calibrate the
reagent pump, if necessary.
1. Access the Pump Flow Rate Test from the Menu screen:
a. Select
3 Troubleshooting and press Enter.
b. Select
1 Fluidics System and press Enter.
c. Select
2 Pump Flow Rate and press Enter.
The Pump Flow Rate Test screen appears as shown in Figure 4-3.
The flow rate units are mL/sec.
Check here for results.
2. Press Start Test.