Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
System Administration 5Ć33
    
Use this procedure to change the time when the system performs the automatic
cleaning of the reagent manifold. Auto Clean occurs once every 24 hours. The
default time is 0200.
On the first day of the month, the system also prints out final statistical summary
reports of the previous month’s QC data at the time of the Auto Clean.
1. Access the System Options screen from the Menu screen:
a. Press
6 System Setup and press Enter.
b. Press 5 Systems Options and press Enter.
The System Options screen appears, as shown in Figure 5-18.
      
    
    

2. Move to the Auto Clean Time field.
3. Type the hour and minutes in the format, HH:MM. You can enter 0 – 23 hours
and 0 – 59 minutes.
4. Press
Done when you finish.
5. Define another setup function or press
Exit Menu to return to the Ready
screen.
  
Use this procedure to:
S define the start date for scheduling maintenance on your system
S increase or decrease the frequency of the maintenance tasks
1. Access the Maintenance Setup screen from the Menu screen:
a. Press
6 System Setup and press Enter.

6
Menu Code
2