Rapidlab 800 Operator's Manual
3. Select pH Only and press Enter.
4. Press
Analyze.
NOTE:ĄĂEnsure that the trailing edge of the sample remains in contact with the
sample ground/temperature sensor.
5. Turn the sample pump clockwise until the leading edge of the sample fills the
reference sensor, as shown in Figure 2-11.
Reference Sensor
Sample Ground/
Temperature Sensor
Sample Pump
6. Press Sample in Place.