Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

   
   
2-POINT CALIBBRATION REPORT APR 12 1994 13:02
SYSTEM 860-1001
Sequence no 433
Calibration Date APR 12 1994
Calibration Time 11:42
CAL POINT
New Meas Drift Units
pH 7.382 7.381 -0.001
pCO
2
35.4 34.9 -0.5 mmHg
pO
2
85.0 85.0 0.0 mmHg
Na
+
140.0 140.9 0.9 mmol/L
K
+
4.00 4.00 0.00 mmol/L
Ca
++
1.25 1.26 0.01 mmol/L
Cl
-
100 99 -1 mmol/L
Glucose 0 0 0 mg/dL
Lactate 0.00 0.02 0 mmol/L
pAtm 755 mmHg
SLOPE POINT
New Meas Drift Units
pH 6.838 6.839 0.001
pCO
2
70.8 71.3 0.5 mmHg
pO
2
0.0 -0.1 -0.1 mmHg
Na
+
100.0 100.3 0.3 mmol/L
K
+
8.00 8.04 0.04 mmol/L
Ca
++
2.50 2.52 0.02 mmol/L
Cl
-
70 70 0.00 mmol/L
Glucose 180 180 0 mg/dL
Lactate 0.00 0.02 0 mmol/L
Atmospheric pressure
appears on all systems
.......
...........
......................
....................
840
850
............................
.................................
.................................
860
.......
...........
..........................
....................
840
850
..............................
.................................
.................................
860
7. Press Done when you finish.
       
more than one report is
found
the Done Options message box appears:
S Select Next Record and press OK to view the next
report that appears on the log.
S Select Previous Record and press OK to view the
previous report that appears on the log.
S Select Search Criteria Screen and press OK to view the
Calibration Data Search Criteria screen.
S Press Cancel to close this message box. The Search
Results screen appears.
one sample is found the Calibration Data Search Criteria screen appears.
If the selected sample has a D code or status message associated with it, the View
Diagnostics F-key appears. Press View Diagnostics to view the Diagnostics
message box.

