NOTE: The default values for the action range are equal to the measurement range.
If you use the default action range values, the system does not flag action range
results but instead indicates that the results are out of measurement range. Change
the action range values to have the system flag action range results.
     
      
pH 7.350 – 7.450* 6.000 – 8.000 840, 850, 860
pCO
2
mmHg 35.0 – 45.0* 5.0 – 250.0 840, 850, 860
pO
2
mmHg 75.0 – 100.0
[
0.0 – 800.0 840, 850, 860
Na
+
mmol/L 135.0 – 148.0
]
70.0 – 200.0 850, 860
K
+
mmol/L 3.50 – 5.30* 0.50 – 20.00 850, 860
Ca
++
1.13 – 1.32
]
0.25 – 5.00 850, 860
Cl
mmol/L 98 – 106* 40 – 160 850, 860
Glucose mg/dL 66 – 93
w
10 – 999 860
Lactate mmol/L 0.5 –2.0
*
w
0.0 – 30.0 860
* Tietz NW ed. Fundamentals of Clinical Chemistry. 3rd ed. Philadelphia: WB Saunders, 1987; 864-891.
[
Weisberg HF. Acid-Base pathophysiology in the neonate and infant. Annals of Clinical and Laboratory
Science 1982; 12(4)249.
]
Lentner C ed. Geigy scientific tables. Vol 3, 8th ed. Basle: Ciba-Geigy Ltd., 1984; 82-83.
w
Toffaletti et al., Clinical Chemistry, 38/12; 2430-2434, 1992
k
Sabata V, Stubbe P, Wolf H. Energy metabolism in the premature fetus. Biology Neonate. 1971; 19:299.
Table 5-8 lists additional default values available on systems with a CO-ox module.
       
     
tHb g/dL 12.0 – 18.0* 2.0 – 27.0
ctO
2
mL/dL 15.0 – 23.0
[
0.0 – 40.0
BO
2
mL/dL 16.0 – 24.0
[
0.0 – 40.0
sO
2
% 92.0 – 98.5 –99.9 – 999.9