3Ć70
Rapidlab 800 Operator's Manual
Hex Tool
OĆring
OĆring
Sample Path
KCI Reservoir
Vent Hole
4. Add KCl fill solution to the KCl reservoir to the fill line, as shown in
Figure 3-42.
4M KCl Fill Solution
Fill Line
OĆring
OĆring
Sample Path
CAUTION: Do not overtighten the reservoir cap. Overtightening can deform the
gasket and cause leaks.
5. Reinstall the reservoir cap and hand-tighten.
6. Continue with Reinstalling a Sensor, page 3-89.
!