Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
Installation HĆ21
S lint-free tissue and swabs
S aspiration adapter
S test/blank sodium or pH sensor (TB2)
S glucose test/blank sensor
(TB4)
S lactate test/blank sensor (TB4)
S reference test/blank sensor (TB5)
S valve wrench
BIOHAZARD: Refer to Appendix A for recommended precautions when working
with biohazardous materials.
     
CAUTION: Exposure to bleach damages the glucose and lactate biosensor
membranes. Replace the glucose and lactate biosensors with the test/blank (TB4)
sensor before cleaning the sample path.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
1. Take the appropriate action.
If you have an . . . Then . . .
840 or 850 Go to step 2.
860 a. Remove the glucose and lactate biosensors.
b. Install the test/blank sensors.
c. Go to step 2.
2. Deproteinize the sample path:
a. Prepare the deproteinizer as directed on the package.
b. Select
2 Maintenance and press Enter.
c. Select 1 Deproteinize and press Enter.
d. Invert the deproteinizer several times to mix.
e. Insert an aspiration adapter into the sample port and insert the other end
into the deproteinizer.
f. Press
Analyze.
g. Remove the adaptor when prompted.
h. Wait 5 minutes for the deproteinizing cycle to finish.
!
860
!
