Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć53
Maintaining the System
9. Press Continue.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
10. Press
Yes to perform a two-point calibration.
   
Materials required:
S 10% solution of household bleach
NOTE: Dilute household bleach (5.25% sodium hypochlorite) 1:10 with reagent
quality water.
S reagent water
S lint-free swabs
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 7 Stop System and press Enter.
2. Remove the required reagent bottle as shown in Figure 3-34.
3. Grasp the reagent fitting and turn it counterclockwise one-quarter turn.
4. Pull the fitting straight out.
      

 
  
  
 
5. Replace the fitting if it is deformed, and then continue with step 8.
WARNING
Wear safety glasses, gloves, and a laboratory coat when handling
bleach.