3Ć76
Rapidlab 800 Operator's Manual
Hex Tool
OĆring
OĆring
Sample Path
KCI Reservoir
Vent Hole
b. Partially fill the KCl reservoir as shown in Figure 3-50.
4M KCl Fill Solution
Fill Line
OĆring
OĆring
c. Tap the front face of the sensor with your knuckle to release any bubbles.
d. Gradually fill the KCl reservoir with KCl fill solution up to the fill line.
CAUTION:ĄĂDo not overtighten the reservoir cap. Overtightening can deform
the gasket and cause leaks.
e. Reinstall the reservoir cap and hand tighten.
10. Continue with Reinstalling a Sensor, page 3-89.
!