Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
4Ć92
800 Series Operator's Manual
    
Use this procedure to disconnect the CO-ox sample path from the base model. You
can continue to measure the remaining parameters.
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1. Turn the CO-ox parameters off:
a. Select
6 System Setup and press Enter.
b. Select 3 Parameters and press Enter.
c. Move to the CO-oximetry frame.
d. Select the tHb parameter and press Enter.
This turns all CO-ox parameters off.
e. Press
Next Screen to access the second screen of parameters.
f. Press Next Screen to access the third screen of parameters.
g. Press Done when you finish.
2. Disconnect the CO-ox and base model sample tubing from the sample
connector.
3. Remove the screw from the sample connector.
4. Invert the sample connector so that the “Y” is on the bottom as shown in
Figure 4-21.
     
Preheater
Sample Connector
Inverted
Connector
6
Menu Code
3