3Ć43Maintaining the System
6. Press Continue.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
7. Press
No.
8. Calibrate the reagent pump from the Menu screen:
a. Select
3 Troubleshooting and press Enter.
b. Select 1 Fluidics System and press Enter.
c. Select 2 Pump Flow Rate and press Enter.
d. Press
Calibrate Pump.
e. Press Exit Test when the Pump calibration complete message appears.
9. Perform a two-point calibration from the Menu screen:
a. Select
1 Calibration and press Enter.
b. Select
2 Two-point and press Enter.
21
Menu Code
3
(from the Main Menu)
1
Menu Code
2
(from the Main Menu)