Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
    
    
Materials required:
S measurement module tubing
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
NOTE: If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
1. Cut a piece of measurement module tubing to the appropriate length as shown
in Figure 3-19.
   
  
840 15.2 cm (6 inches)
850 12 cm (4.75 inches)
860
11.43 cm (4.5 inches)
2. Stop the system from the Menu screen:
a. Select 2 Maintenance and press Enter.
b. Select 7 Stop System and press Enter.
3. Push up the latches on the measurement module door and lift the door.
4. Take the appropriate action.
            
840 step 5.
850 or 860 step 6.
2
Menu Code
7