Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć103
Maintaining the System
8. Pinch the ends of the clamp and pull it away from the bracket.
9. Remove the old lamp from the clamp and discard it.
CAUTION: Avoid touching the lamp with your fingers.
10. Place a new lamp in the clamp with the front of the lamp facing the system.
11. Pinch the ends of the clamp together and reinstall it in the bracket.
12. Reinstall the connector in the system.
13. Reinstall the lamp cover by inserting the tabs on the cover into the system and
snapping the cover back into place.
14. Reconnect the power cord to the power source and allow the system to warm
up for at least 15 minutes.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
15. Press
No.
16. Verify the system component temperatures from the Menu screen:
a. Select
3 Troubleshooting and press Enter.
b. Select 2 Temp/pAtm and press Enter.
c. Press Start Test.
d. Verify that the system components warm up.
e. Press
Stop Test.
f. If the temperature control system is off, press Reset Control and allow the
system to warm up.
g. Press Exit Test.
17. When the temperature is stable, perform a two-point calibration from the
Menu screen:
a. Select
1 Calibration and press Enter.
b. Select 2 Two-point and press Enter.
Replacing the COĆox Lamp
WARNING
Burn Hazard. Do not touch or look at the lamp when the lamp is
illuminated. Injury to the skin or the eye can occur when the lamp is operated
outside the lamp housing.
Materials required:
S CO-ox lamp
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
!
3
Menu Code
2
1
Menu Code
2
(from the Main Menu)
2
Menu Code
7