Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć49
Maintaining the System
    
Materials required:
S 0.022 inch diameter clot-removal line (to fit through a 0.030 inch ID pathway)
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select
7 Stop System and press Enter.
2. Push up the latches on the measurement module door and lift the door.
3. Push the spring-loaded latch to the right as shown in Figure 3-31.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
4. Grasp the tab on the sample ground/temperature sensor and pull the sensor up
and out of the measurement module.
       
  
   
      
        
 
5. Push a clot removal line through the sensor to clear away any obstructions as
shown in Figure 3-32.
    
  
