Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
System Administration 5Ć59
  
        
pCO
2
2.00 – 7.90% 5.00% (Cal CO
2
+ 2.9) –
99.9%
10.00%
pO
2
5.00 – 98.00% 12.00% 0.00% (fixed) 0.00%
1. Access the Calibration Gas Values screen from the Menu screen:
a. Select 5 Operating Setup and press Enter.
b. Select
6 Calibration Setup and press Enter.
c. Select 3 Gas Values and press Enter.
The Calibration Gas Values screen appears, as shown in Figure 5-36.
    
  


   
    
2. Move the cursor to the field that you want to change.
3. Type the new gas value and press
Enter.
4. Repeat steps 2 and 3 for each gas value you want to change.
5. Press
Done.
6. Define another setup function or press
Exit Menu to return to the Ready
screen.
