800 Series Operator's Manual5Ć50
       
  
   
 
4. Perform one of the following options.
          
change to a new
QC lot
type the new QC file information in the fields or, if you use
the optional bar code scanner, scan the new QC file
information. Continue with step 5.
edit the ranges for
this QC file
continue with step 5.
edit fields on this
screen
move to the field requiring a change and type the new data.
Press
Enter after you complete each field. Continue with
step 5.
5. Press Done.
A message appears prompting you to delete the current QC file setup data.
NOTE:ĄĂYou must delete the current QC file data when you change lots.
Otherwise, the QC file will contain data from the old and the new lots.
         
changing to a new
QC lot
press Yes. The system prints a copy of the current QC file
setup data, deletes this data, and prints a copy of the QC
statistical summary report.
The QC File Setup (ranges) screen appears with all data fields
empty, as shown in Figure 5-30. Continue with step 6.
editing the existing
QC file ranges
press No. The QC File Setup (ranges) screen appears. The
fields contain the data for the file you selected. Continue with
step 6.