Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
Installation HĆ25
NOTE: Do not insert swabs into the sample port or spray anything into the
measurement module.
2. Rinse the exterior surfaces with reagent water.
3. Clean spills around any of the roller cages, if required:
a. Remove the roller cage as described in Replacing a Roller Cage in
Section 3.
b. Clean the roller cage and the roller cage shaft with a 10% solution of
household bleach.
c. Rinse with reagent water and dry thoroughly.
d. Reinstall the roller cage.
4. Empty and discard the waste bottle. Clean waste outlets and waste outlet cover
as described in Emptying the Waste Bottle in Section 3.
5. Remove the printer paper. Refer to Replacing the Printer Paper in Section 3.
6. Tape the printer cover in place.
   
1. Shut down the system from the Menu screen:
a. Select
7 System Utilities and press Enter.
b. Select 3 Shutdown and press Enter.
c. Press Yes.
The Shutdown screen appears.
CAUTION:ĄĂDo not unplug the system until the following message appears on
the screen:
...synching disks... done
This is an operating system message, indicating that the system has
successfully completed the shutdown procedures. Unplugging the system
before this message appears can damage the system.
2. When you see the operating system message, you can disconnect the power
cord from the power supply.

!