1. Access the Select Default Panels screen from the Menu screen:
a. Select
6 System Setup and press Enter.
b. Select
4 Panels and press Enter.
c. Select
1 Default Panel and press Enter.
The Select Default Panel screen appears, as shown in Figure 5-4.
        
   
   
2. Select the appropriate panel and press Enter.
Custom panels are listed as Panels 1–5. Custom panels are defined during
system setup by the system administrator or authorized personnel. Refer to
Defining Custom Panels on page 5-9.
3. Press
Done when you finish.
   
Use this procedure to select the default set of parameters you want the system to
analyze. If you select a parameter to be off, the system does not measure or
calculate results during analysis, and the parameter name does not appear on the
Results screen or the printed report. The default value for all parameters is on,
except for O
2
SAT, O
2
CT, and [Qsp/Qt (est,T)].
NOTE: Any calculated parameters that rely on another parameter value, which has
been selected to be off, are not calculated. However, the parameter name appears on
the Results screen, but no value is reported and the parameter name and value do
not appear on the printed report.
1. Access the Parameters screen from the Menu screen:
a. Select
6 System Setup and press Enter.
b. Select
3 Parameters and press Enter.
The Parameters screen appears, as shown in Figure 5-5.

