Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
Installation HĆ23
     
1. Remove all the reagent bottles from the system and perform a prime sequence.
2. Access the Prime screen from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select 3 Prime and press Enter.
3. Select
All Reagents and press Enter.
4. Press
Done.
5. Press
Menu to return to the Menu screen.
6. Press
No.
7. Access Fluidics Functions from the Menu screen:
a. Select
3 Troubleshooting and press Enter.
b. Select
1 Fluidics System and press Enter.
c. Select 1 Fluidics Functions and press Enter.
8. Select
C1/C2 and press Enter.
9. Press
Start Test.
10. Press
Exit Test.
11. Empty, rinse, and fill the four reagent bottles with reagent water.
12. Install the reagent bottles containing reagent water in place on the system.
13. Perform a Prime sequence as described in steps 2 through 6.
14. Perform an C1/C2 Fluidics Function Test as described sequence 7 through 10.
15. Remove the reagent bottles.
16. Air dry the pathways:
a. Repeat steps 2 through 6 without the reagent bottles in place.
b. Repeat steps 7 through 10 without the reagent bottles in place.
   
1. Stop the system from the Menu screen:
a. Select 2 Maintenance and press Enter.
b. Select
7 Stop System and press Enter.
2. Grasp the sample door and pull it to the right as shown in Figure H-6.
3. Grasp the tab on the retainer ring and firmly pull the tab toward you to rotate
the ring.


