6. Clean the fitting with a swab moistened with a 10% solution of household
bleach.
7. Rinse the fitting with reagent water and dry thoroughly.
Ensure that the O-ring is in place on the back of the reagent fitting.
8. Reinstall the fitting into the appropriate position on the reagent manifold, and
turn clockwise one-quarter turn.
Ensure that the fitting fits tightly into the reagent manifold.
9. Reinstall the reagent bottle.
Push the bottle to ensure that it fits tightly on the reagent fitting.
10. Repeat steps 2 through 9 to remove and clean other reagent fittings as
required.
11. Press
Continue.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
12. Press
No.
13. Perform a prime sequence:
a. Select
3 Prime and press Enter.
b. Select the appropriate reagents to prime and press Enter.
c. Press Done.
d. Press
Home or Menu when the Prime menu appears.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
14. Take the appropriate action.
If . . . Then . . .
you want to perform a two-point
calibration
press Yes.
you do not want to perform a
two-point calibration
a. Press No.
b. Press
Home to return to the Ready screen.