Use this procedure to calibrate the tHb slope for the CO-ox module. You can
perform this calibration from either the Analyze mode or
the Menu mode.
1. Initiate the tHb slope calibration procedure.
Analyze mode a. Press Calibrate.
b. Select
tHb Slope.
Menu mode a. Select 1 Calibration and press Enter.
b. Select
9 tHb Slope and press Enter.
The tHb Slope Calibration screen appears, as shown in Figure 2-33.
2. Enter the slope value for the slope reagent and press Enter.
NOTE:ĄĂThe last-entered value is displayed. To remove this value, press the
Clear Entry F-key and enter the new value.
3. Press
Done.
4. Insert the sample device into the sample port.
5. Press
Analyze.
6. When prompted, remove the sample device.
The system performs a wash at the end of the calibration and then returns to
the Ready screen.