Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

  
The following result flags can appear on printed reports:
This flag . . . Indicates . . .
* the measurement did not reach endpoint.
the calibration is above the upper drift limit.
the calibration is below the lower drift limit.
- - - the result is above the upper limit of the measurement range.
- - -
the result is below the lower limit of the measurement range.
The following table describes the messages that may appear in calibration reports
on the roll printer:
Message Description
or = D2
Excessive Drift: __
Sensor drift is beyond predefined limits during a one-point or a
two-point calibration.
* = D5 No Endpoint:
__
Calibration did not reach endpoint.
D3 Slope Error: __ Sensor slope is beyond predefined limits during a two-point
calibration.
D4 Offset Error: __ Sensor offset is beyond predefined limits during a one-point or
a two-point calibration.
D22 Barometric
Pressure Error
(for gas cals)
Signal from the barometer detects atmospheric pressure beyond
predefined limits.
D50 Glucose Sensor
Error
System detects an open connection in the glucose sensor.
D51 Lactate Sensor
Error
System detects an open connection in the lactate sensor.