Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
3Ć48 Rapidlab 800 Operator's Manual
     

 


 
 

   
   
6. Tap the front face of the sensor with your knuckle to release any bubbles.
7. Clean any salt deposits on the reference sensor with a lint-free swab moistened
with deionized water and dry it thoroughly.
8. Ensure that the O-rings are in place on both sides of the sensor.
9. Clean the O-ring area on the spring-loaded latch with a lint-free swab
moistened with reagent water.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
10. Reinstall the sensor:
a. Gently align the top of the reference sensor with the sensor contact.
b. Snap the sensor down into place.
c. Press the tab on the spring-loaded latch to release the latch.
11. Close the measurement module door.
12. Press
Continue.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
13. Allow the system to warm for up at least 15 minutes.
14. When the temperature is stable, press
Yes to perform a two-point calibration.
After the sensor temperature equilibrates, remove the sensor and inspect for
bubbles. As the temperature of the sensor rises to 37°C, gas is driven from the
solution, causing bubbles. Remove any bubbles present.
!

