Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 

  
System Message Description
No Sample Device
Detected
Appears when you press Analyze, the sample door closes, and no
device is detected in the sample port. See Troubleshooting System
Messages, page 4-120.
No Waste Bottle
Detected
Appears when the waste bottle is not installed.
__ Not Sent Appears when you press Do Not Send at the end of analysis.
Out of Range Appears when results for a parameter are out of the measurement
range.
Probe Detected
Obstruction
Appears when the probe hits an obstruction such as the plunger
because there is insufficient sample.
Report Results with
Cal Drift
Appears when results are reported for a sensor that exhibits
calibration drift.
Sample
Temperature
Out of Range
Appears when the measurement block temperature is not in range
at the end of measurement sequence. See Troubleshooting System
Messages, page 4-120.
__ Sent Appears when you press Send at the end of analysis or if results
are sent automatically.
SulfHb > 1.5% Appears when the CO-ox module detects a concentration of
SulfHb greater than 1.5%.
tHb Slope
Calibration Due
Appears when you need to perform a tHb Slope calibration for
the CO-ox module.
Unexpected Device
Detected
Appears if the system detects an unexpected device, such as
when a sample device is inserted as the door attempts to close to
perform an automatic sequence.
Waste Full Appears when the waste bottle can no longer accept waste.
Waste Is Almost
Full
Appears when the waste bottle is near capacity and can accept
waste from three more samples.