NOTE:ĄĂThe system uses two sequence numbers for two-point calibrations.
Only one number appears in the log.
5. Perform the appropriate task at the Calibration Data Search Log.
view the data for a
calibration
select the report you want and press Enter. The
Calibration Data Search Result screen appears, as shown
in Figure 2-35.
print a calibration report a. Select the report you want.
b. Press
Reporting Options.
c. Press
OK. The calibration report prints, as shown in
Figure 2-36.
transmit the results to an
LIS or a data management
system
a. Select the report you want.
b. Press
Reporting Options.
c. Select Send Results and press
Enter.
d. Press
OK.