System messages provide information about the operating status of the system and
can appear in the status area of the screen, in the Status Event Log, or on printed
reports. Table 4-1 describes each system message.
 
  
1-pt Cal Pending
in __ Min
Appears 5 minutes before a one-point calibration is due and
counts down until the calibration starts.
1-pt Metabolite Cal
Due
Appears during an analysis to indicate the system will perform a
metabolite calibration at the completion of the analysis.
1-pt Metabolite Cal
Due in __ Min
Appears 5 minutes before a one-point metabolite calibration is
due and counts down until the calibration starts.
2-pt Cal Pending
in __ Min
Appears 5 minutes before a two-point calibration is due and
counts down until the calibration starts.
Additional
Messages
Appears in the status area to indicate that there are more
messages, which can be viewed in the status log.
Auto Clean
Pending in __ Min
Appears 5 minutes before an Auto Clean sequence is due and
counts down until the sequence starts.
Bubbles Detected
In Sample
Appears when the system detects a non-continuous fluid in the
measurement module sample path. See Troubleshooting System
Messages, page 4-114.
Calibration
Overdue: __
Appears when the system must perform a calibration on the
sensor(s) indicated.
COox Cover Open
During Meas
Appears when the cover of the CO-ox module is open during a
sample or slope measurement. Question the results.
COox Cover Open
During Zero
Appears when the CO-ox module cover is open during the CO-ox
module zero measurement sequence of a one- or two-point
calibration. Repeat the calibration with the CO-ox cover closed.
COox Cover is
Open
Appears when the cover of the CO-ox module is open when you
begin the sample measurement. You must close the cover before
pressing Analyze to ensure an accurate CO-ox analysis.
(Continued)