b. Select 3 Measurement and press Enter.
c. Select 2 COox Optics and press Enter.
The COox Optics Test screen appears.
13. Press
Start Test.
14. Check the screen for the message, Lamp test passed.
15. Press
Exit Test.
16. Press
Exit Menu.
17. Perform a one-point calibration from the Menu screen:
a. Select 1 Calibration and press Enter.
b. Select
1 One-point and press Enter.
   
Materials required:
S 2 fuses of the appropriate rating
S small, flat-blade screwdriver
Refer to Table 3-2 to identify the correct fuses for the voltage you use.
  
    
100/120V 4A Slo Blo 5 x 20 mm
220/240V 2A Slo Blo 5 x 20 mm
WARNING
To prevent electrical shock or damage to the system, remove power
from your system before performing this procedure.
1. Shut down the system from the Menu screen:
a. Select
7 System Utilities and press Enter.
b. Select 3 Shutdown and press Enter.
c. Press
Yes.
CAUTION: You must wait at least 1 minute before you disconnect the power cord,
and then wait at least 10 seconds before you reconnect the power cord. If you do
not adhere to the time intervals, you can damage the system.
2. Wait until prompted and then disconnect the power cord from the power
source.
1
Menu Code
1
(from the Main Menu)
7
Menu Code
3
!