Maintaining the System
Materials required:
S sample probe
BIOHAZARD: Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1. Stop the system from the Menu screen:
a. Select
2 Maintenance and press Enter.
b. Select
7 Stop System and press Enter.
2. Remove the old probe:
a. Disconnect the sample tubing from the connector on the measurement
module.
b. Disconnect the sample tubing from the bracket on the probe mount.
c. Rotate the sample tubing toward you by pulling it up until it is
perpendicular to the bench top as shown in Figure 3-67.
Sample Tubing
Sample Probe
Bracket
Probe Mount
d. Disconnect the sample tubing from the sample probe.
e. Pull the probe tab to the left into the slot, and pull the probe out of the
sample port, as shown in Figure 3-68.