Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
EĆ28 800 Series Operator's Manual
   
All performance data presented in this section was generated using 860 systems.
The system used default correlation factors, and performed calibrations using the
default settings recommended by Bayer Diagnostics for optimum performance. All
reported values were corrected to 760 mmHg. The operating environment during
the collection of this data was normal room temperature (about 23_C).
You should determine your own performance characteristics in your laboratory
with your 860 system.
   
To test for interferences, serum was spiked with a potentially interfering substance
up to the test concentration shown in the following two tables. The interference was
calculated by comparing the spiked sample to an unspiked sample immediately
preceding it. To evaluate the interference from fluoride/oxalate, the blood was
drawn in grey-top tubes and compared to the glucose recovery in heparinized
blood.
Table E-23 lists substances that were found not to interfere with the glucose
measurement. At the concentrations listed, these compounds produced less than
6 mg/dL (0.3 mmol/L) error in the recovered glucose concentration.
   
  
Chlorpromazine 5 mg/dL
Dopamine 0.5 mg/dL*
Ethanol 350 mg/dL
Salicylate 50 mg/dL
Sodium Nitroprusside 70 mg/dL
Citrate 1000 U/dL
Heparin 20,000 U/dL
Acetoacetate 40 mg/dL
Ascorbate
6 mg/dL
(Continued)