Bayer HealthCare Rapidlab 800 Blood Glucose Meter User Manual


 
Performance Characteristics EĆ17
   
All performance data presented in this section was generated using 850 systems.
The system used default correlation factors, and performed calibrations using the
default settings recommended by Bayer Diagnostics for optimum performance. All
reported values were corrected to 760 mmHg. The operating environment during
the collection of this data was normal room temperature (about 23_C).
You should determine your own performance characteristics in your laboratory
with your 850 system.
  
Quality control materials and calibration verification materials were analyzed on
the 850 systems. The results are presented here.
Precision on aqueous quality control materials was estimated using four 850
systems. At least seven runs per instrument were made over ten days. Two
replicates of each control level were analyzed in each run.
Table E-14 summarizes the results of the 850 system precision on QC materials.
   
   * 
[
pH 1 72 7.169 0.002 0.002
2 71 7.428 0.001 0.003
3 72 7.623 0.001 0.002
pCO
2
1 72 71.5 0.43 1.01
2 71 44.1 0.21 0.58
3 72 22.6 0.13 0.38
pO
2
1 72 58.3 2.64 3.25
2 71 98.2 1.93 3.32
3 72 142.2 1.90 2.92
Na
+
1 72 113.1 0.20 0.81
(Continued)