6. After the system copies the software files, the Install Identification message
box appears.
If . . . Then . . .
you have more program
diskettes to install
the Insert Next Diskette message appears on the
Install Identification message box instruction line.
NOTE: Do not press Continue until you have inserted
the diskette.
a. Remove the diskette from the diskette drive.
b. Insert the next program diskette.
c. Repeat steps 4 and 5.
the Last Diskette message
appears on the Install
Identification message
box instruction line
a. Press OK.
A message appears prompting you to remove the
diskette.
b. Remove the diskette from the diskette drive.
The Processing Software Installation message
appears. The system emits a beep every few
seconds until the installation is complete.
NOTE:ĄĂIf the Last Diskette message does not appear after the system reads
the last program diskette, it is likely that at least one of the program diskettes
was not installed. Cancel the installation and start the procedure again.
When the system finishes the software installation, the Install Software
Finished screen appears prompting you to shut down the system.
7. Press
OK to shut down the system.
The system initiates an automatic shut down.
A message appears on the screen and on the roll printer, directing you to wait
before you disconnect the power.
CAUTION:ĄĂDo not unplug the system until the following message appears on
the screen:
...synching disks... done
This is an operating system message, indicating that the system has
successfully completed the shutdown procedures. Unplugging the system
before this message appears can damage the system.
8. When you see the operating system message, disconnect the power cord from
the power supply.
CAUTION:ĄĂWait at least 10 seconds before you reconnect the power cord. If
you do not adhere to the time intervals, you can damage the system.
!
!