9. Press Yes to perform a two-point calibration.
10. Analyze a minimum of two levels of quality control material to verify sensor
performance.
Press
CANCEL to stop conditioning. The system performs a wash and then
displays the Calibrate System message box.
    
The main tank pressure must be higher than 300 psi. The second stage pressure
must be 3 to 5 psi.
1. Check the main tank and second-stage pressure for each gas tank by looking at
pressure gauges on the gas tank regulator as shown in Figure 3-4.
   
 
 

 
 
  

2. If the main tank pressure is below 300 psi, verify that the gas tank is open.
3. If the main tank is open and the pressure is still below 300 psi, replace the tank
as described in Replacing the Gas Tanks, page 3-90
.

