D35 Electronics Error:
Problem
The system detects an error in the electronics system.
Qualifiers: 1 2 4 5 6 7 8 9 10 11 12 13
Solutions
Qualifier Procedure
1, 2, 4, 5, 8,
9, 11
perform the Measurement Test as described on page 4-70.
If the measurement test does not remove the D35, shut the system
down and restart it as described in Shutting Down and Restarting the
System in Section 5.
6 or 7 shut down and restart the system as described in Shutting Down and
Restarting the System in Section 5.
10 check the temperature control as described in Temperature/pAtm Test,
page 4-69.
12 perform the Valves Test as described on page 4-62.
13
perform the Pump Functions Test as described on page 4-58.
If the D code reappears, contact your Service Representative.
D35